The Smartmeasurement Differential Pressure Wedge Meters (ALDPWG) represents one of the many DP flow technologies available in our program. The ALDPWG meters consist of a flow body with (see below) wedge “V” shape throttle restriction. This restriction creates a differential pressure between ((P1) before and (P2) after the wedge (P2))-see below which is proportion to the square of the volumetric flow rate. The wedge restriction has no critical surface dimensions, or sharp edges, that will affect measurement accuracy as the result of normally expected wear. The ALDPWG meter is excellent for very low flows or low Reynolds number (Re. < 500), high-viscosity fluids which contains dust or solid particle or applications such as liquid cement, crude oil coal-tar oil, wastewater, hard pulp, asphaltic oil, coal-tar oil, wastewater, hard pulp, carbon black solution, two phase flows, etc. The ALDPWG meter must accompany a Differential Pressure (DP) transmitter to indicate, rate, total etc. The ALDPWG can be used with our ALDP3051 or ALDP1051 or customers own DP transmitters.
Catalog
- Measures Liquids, Gas and Steam
- Bi-directional flow
- Low Reynolds number (<500)
- Minimum upstream and downstream requirements
- Can measure clean or dirty liquids, slurries, gases and steam
- Standard sizes 0.5” to 24”(15~DN600 mm)
- Can be supplied with ALDP3051 DP Transmitter
- High temperature application with remote mount DP transmitter
- Good particle tolerance; not sensitive to suspended contaminants
| Sizes |
0.5" to 24"(15~DN600 mm) |
| Measuring Range |
Re < 500 |
| Materials |
Body material: Cast lron, stainless steel, high temperature alloy.
Wedge: Stainless steel, Monel, Hastelloy C, high temperature alloy. |
| Accuracy |
(calibrated): 1%, 0.5%
(un-calibrated): 3% |
| Repeatability |
±0.1% of reading |
|
|
| Turn down ratio |
Better than 10:1 |
| Process Connections |
ANSI and DIN flanges |
| Transmitter Connections |
3 in. RF flange (SCH 80 I.D.) |
| Temperature |
up to 750 °F (400℃) |
| Pressure |
6000 PSIG (42.0MPa) |
| Transmitter Requirements |
ALDP 3051 transmitter |
|